Appendix MultiExact Gas Analyser
A9.4 Performance data: 0-10ppm CO
2
measurement
The display indication given below is the default indication. You can configure
the analyser to provide other display indications (see Section 5.10).
Performance data has been determined in accordance with EN61207.
Display indication
ppm CO
2
Full scale range
0 to 10 ppm CO
2
Analogue output range
User selectable
(minimum difference 10% of full range)
Resolution
0.01 ppm CO
2
Repeatability
< 0.1 ppm CO
2
Linearity
± 0.1 ppm CO
2
Intrinsic error (accuracy)
± 0.1 ppm CO
2
Drift per week
± 0.2 ppm ± 2% of reading
Output fluctuation
± 0.1 ppm CO
2
Response time *
15 seconds
Flow effect
Flow driven
± 0.1 ppm or 1% of reading **
Pressure driven
± 0.1 ppm or 1% of reading **
Temperature coefficient
± 3% of measurement or 0.25 ppm per 10 °C **
Pressure effects
0.4% of measurement
#
*
T
90
at 500 ml min
-1
.
For a 200 to 500 ml min
-1
change in sample gas flow rate.
For a 14 to 55 kPa (0.14 to 0.55 bar, 2 to 8 psi) change in sample gas inlet pressure.
#
For a 1% change in pressure.
** Whichever is the greater.
128 05410001A / Revision 6

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Servomex SERVOPRO MultiExact and is the answer not in the manual?

Servomex SERVOPRO MultiExact Specifications

General IconGeneral
BrandServomex
ModelSERVOPRO MultiExact
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals