EasyManua.ls Logo

Servomex SERVOPRO MultiExact - A9.9 Performance data: TCD

Servomex SERVOPRO MultiExact
145 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Appendix MultiExact Gas Analyser
A9.4 Performance data: 0-10ppm CO
2
measurement
The display indication given below is the default indication. You can configure
the analyser to provide other display indications (see Section 5.10).
Performance data has been determined in accordance with EN61207.
Display indication
ppm CO
2
Full scale range
0 to 10 ppm CO
2
Analogue output range
User selectable
(minimum difference 10% of full range)
Resolution
0.01 ppm CO
2
Repeatability
< 0.1 ppm CO
2
Linearity
± 0.1 ppm CO
2
Intrinsic error (accuracy)
± 0.1 ppm CO
2
Drift per week
± 0.2 ppm ± 2% of reading
Output fluctuation
± 0.1 ppm CO
2
Response time *
15 seconds
Flow effect
Flow driven
± 0.1 ppm or 1% of reading **
Pressure driven
± 0.1 ppm or 1% of reading **
Temperature coefficient
± 3% of measurement or 0.25 ppm per 10 °C **
Pressure effects
0.4% of measurement
#
*
T
90
at 500 ml min
-1
.
For a 200 to 500 ml min
-1
change in sample gas flow rate.
For a 14 to 55 kPa (0.14 to 0.55 bar, 2 to 8 psi) change in sample gas inlet pressure.
#
For a 1% change in pressure.
** Whichever is the greater.
128 05410001A / Revision 6

Table of Contents

Related product manuals