MultiExact Gas Analyser Appendix
A9.6 Performance data: 0-10ppm CO measurement
The
display indication given below is the default indication. You can configure
the analyser to provide other display indications (see Section 5.10).
Performance data has been determined in accordance with EN61207.
Display indication
ppm CO
Full scale range
0 to 10 ppm CO
Analogue output range
User selectable
(minimum difference 10% of full range)
Resolution
0.01 ppm CO
Repeatability
< 0.1 ppm CO
Linearity
± 0.1 ppm CO
Intrinsic error (accuracy)
± 0.1 ppm CO
Drift per week
± 1 ppm ± 1% of reading
Output fluctuation
± 0.1 ppm CO
Response time *
20 seconds
Flow effect
Flow driven
± 0.1 ppm or 1% of reading **
Pressure driven
± 0.1 ppm or 1% of reading **
Temperature coefficient
± 3% of measurement or 0.3 ppm per 10 °C **
Pressure effects
< 1% of measurement
#
*
T
90
for change of 0-5ppm at 400 ml min
-1
.
For a 200 to 500 ml min
-1
change in sample gas flow rate.
For a 14 to 55 kPa (0.14 to 0.55 bar, 2 to 8 psi) change in sample gas inlet pressure.
#
For a 1% change in pressure.
** Whichever is the greater.
05410001A / Revision 6 131

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Servomex SERVOPRO MultiExact and is the answer not in the manual?

Servomex SERVOPRO MultiExact Specifications

General IconGeneral
BrandServomex
ModelSERVOPRO MultiExact
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals