MultiExact Gas Analyser 5 – Installation and set-up
6. Scroll down to the high target value
then repeat steps 3 and 4 to set the
high target and high tolerance
values.
Figure 33 – The auto val/cal
high target value screen
Figure 34 – The auto val/cal
high tolerance screen
5.5.5 Automatic validation/calibration type and mode
1. Return to the auto val screen
(Figure 30). Select 'Control'.
2.
Select required type: 'Low', 'High',
'Low & High', 'High & Low'.
Figure 35 – The auto val/cal
type screen
3. Select 'Mode'. Select required type:
'Validate' or 'Calibrate'.
Figure 36 – The auto val/cal
mode screen
5.5.6 Automatic validation/calibration phases
1. Return to the auto val screen
(Figure 30). Select 'Phase'.
2. Select 'On' or 'Off' for Pre warning.
Figure 37 – The auto val/cal
phase screen
05410001A / Revision 6 47