Appendix MultiExact Gas Analyser
A9.9 Performance data: TCD
The display indication given below is the default indication. You can configure
the analyser to provide other display indications (see Section 5.10).
Performance data has been determined in accordance with EN61207.
Display indication
%
Analogue output range
Resolution
0.01%
Repeatability
0.5% of range
Linearity
± 1% of range
Intrinsic error (accuracy)
± 1% of range
Drift per month
± 1% of range
Output fluctuation
± 0.5% of range
Response time *
T90 15 seconds *
Flow effect
< 0.1%
#
Temperature coefficient
± 1% of range per 10 °C
*
for 150 ml min
-1
.
Flow driven.100
200 ml min
-1
or
pressure driven 2
– 8 psig.
#
Over defined flow range
Gas and Ranges available
0-1%
0-2%
0-5%
0-10%
0-20%
0-30%
0-50%
0-100%
80-100%
90-100%
Ar in N
2
Ar in O
2
N
2
in Ar
He in N
2
He in O
2