EasyManuals Logo

Servomex SERVOPRO MultiExact User Manual

Servomex SERVOPRO MultiExact
145 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #65 background imageLoading...
Page #65 background image
MultiExact Gas Analyser 5 – Installation and set-up
When you select
display units other than the measurement default units, you
must also enter the units conversion factor: refer to Appendix A5
to determine
the units conversion factor for your specific application.
If you select the "off"
option on the units selection screen, the display units
revert to the measurement default units as supplied.
Converting from one measurement unit to a different display measurement
unit may reduce the resolution of the displayed measurements.
Use the following procedure to select the displayed units, and to change the units
conversion factor:
1. Select
Set up
Unit selection.
2. If you want to change the currently
displayed units, press the soft
key: the units selection screen will
then be displayed.
Figure 60The currently selected
units screen
3. If required scroll to the 'Factor' screen
to change the unit conversion factor.
Figure 61The units selection
screen
4. Edit the displayed offset as described
in Section 4.10
referring to
Appendix A5 for the correct value.
Figure 62The units conversion
factor screen
05410001A / Revision 6 59

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Servomex SERVOPRO MultiExact and is the answer not in the manual?

Servomex SERVOPRO MultiExact Specifications

General IconGeneral
BrandServomex
ModelSERVOPRO MultiExact
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals