EasyManuals Logo

Shimadzu ATX-R Series User Manual

Shimadzu ATX-R Series
160 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #127 background imageLoading...
Page #127 background image
126
V
Connecting External Equipment
10
CONNECTION AND COMMUNICATION WITH PERIPHERAL DEVICES
Commands relating to calibration
Command Function
ECAL Starts external calibration
ECAL.W = XXX.XXXX
Sets the reference weight value (W ref) for
calibration
XXX.XXXX: Setting value
ICAL Executes calibration with the internal weight
Commands relating to zero / taring
Command Function
ZRNG = X.XXXX
*1
Sets the zero range X.XXXX: Setting value
Commands relating to unit registration
Command Function
g Sets gram units
mg Sets milligram units (only accepted by models capable of displaying 0.001 g)
ct Sets carat units
mom
*1
Sets momme units
Other companies' commands
Command Function
TI Immediate taring (Mettler)
S Single output at stability (Mettler)
SI
*1
Immediate single output (Mettler)
SIR
*1
Continuous output (Mettler)
SR
*1
Continuous output at stability (Mettler)
(ESC) P Immediate single output (Sartorius) ESC = &H1B
(ESC) T Immediate taring (Sartorius) ESC = &H1B
Others
Command Function
"_" (space) Buffer clear command
# = XXXXXXX Enters and displays a numerical value XXXXXXX : Numerical value
{oo ...
Echo-back mode
oo ...: Character string
*1 Not applicable to a veried balance as a legal measuring instrument.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Shimadzu ATX-R Series and is the answer not in the manual?

Shimadzu ATX-R Series Specifications

General IconGeneral
BrandShimadzu
ModelATX-R Series
CategoryScales
LanguageEnglish

Related product manuals