BEFORE WEIGHING
This section will explain processes from balance installation to the beginning of measurement.
Choosing the Installation Site
The measuring performance of the balance is greatly inuenced by the environment where it is
installed.
Observe the following points to ensure safe and accurate weighing.
! Caution
Prohibitions
Do not use the balance anywhere exposed to explosive, combustible
or corrosive gases.
This could cause re or trouble.
Instructions
Use the correct power supply and voltage with the balance.
Use the balance with the attached AC adapter.
Using an incorrect power supply or voltage with the balance will lead to re or trouble
with the balance.
Note also that if the power supply or voltage is unstable or if the power supply capacity
is insufcient, it will not be possible to obtain satisfactory performance from the
balance.
Precautions on Use
Prohibitions
Avoid locations where the balance will be exposed to any of the
following.
You may not be able to obtain correct weight readings.
•
Air ow from an air conditioner, ventilator, door or window
•
Extreme temperature changes
•
Vibration from surroundings or nearby equipment
•
Direct sunlight
•
Dust, ne particles, electromagnetic waves or a magnetic eld
Instructions
Install the balance on a strong and stable at table or oor in the
room.
Placing the balance in an unstable site could lead to injury or trouble with the balance.
When selecting the installation site, take into account the combined weight of the
balance and the item to be weighed.
Installation