58
4
CALIBRATION
In the ATX-R series, the weight for calibration is built in. The internal weight itself is calibrated on
shipment from the factory, but it is possible to recalibrate it using external weights. This is called
P.CAL
For the range of values for the external weights that can be used, refer to "Specications" (^ page 146).
*
Not applicable to a veried balance as a legal measuring instrument.
1
Press for about 3 seconds.
This opens the calibration menu.
2
Select calibration of the internal weight.
[P.CAL]
3
Enter the administrator’s password.
^
"Entering Numerical Values", page 45
^
"Changing the Password", page 142
If the password is wrong…
The error message shown to the right will be
displayed and the display will return to the
indication in step 1.
4
Press
The password will be recognized and the display
will blink zero. The zero point will be read, so
place no loads on the pan and wait.
*
If "WAIT" is displayed…
The calibration record is being output.
When output has nished, span
calibration will start automatically.
After a short time the external sensitivity
calibration weight value will blink.
!P!-!0!0!0!0
/
/
/
/
/
/
/
/
/
/
/
/
/
/
! !E!R!R! !N
W!A!I!T!
! !0.0!0!0!0
2!0!0.0!0!0!0
/
/
/
/
/
/
/
/
/
/
/
/
/
/
/
/
/
/
/
/
/
/
/
/
/
/
/
/
/
/
/
/
/
/
/
/
/
/
/
/
/
/
/
/
/
/
/
/
/
/
/
/
/
/
/
/
/
/
/
/
/
/
/
/
/
/
/
/
/
/
/
/
Calibration of the
Internal Weight
(ATX- R Only)