Auto zero range value
The auto zero range value is only
effective in the units that are displayed
when the value is entered.
If other units are later selected, change
(update) the setting for the zero range
value by following the procedure
from step
1
while these new units are
displayed.
The upper limit value for the zero range
is 99 d. 1 d is the minimum indication in
the displayed units.
For example, for a balance with a
minimum indication of 0.0001 g, the
situation is as follows.
Units
Minimum
Indication
Upper Limit
Value for the
Zero Range
g 0.0001 g 0.0099 g
ct 0.001 ct 0.099 ct
4
Return to the weighing mode.