113
5.1 Trouble shooting
5
*1 If the jet is damaged; it needs to be replaced. If the collector is contaminated with white powder, it needs to be cleaned.
In both cases, contact your Shimadzu representative.
A lot of noise or noise
level is high.
Carrier gas quality is poor. Replace carrier gas with a purer
one.
Provide a molecular sieve filter
in carrier gas flow line.
Hydrogen gas quality is poor. Replace hydrogen gas with a
purer one.
The quartz jet is dirty. *1 Clean dust on the jet by
spraying air from above.
Problem Possible cause Solution Reference