GCMS-QP2010 Series Operation Guide 21
4.1 Creating a Method File
4
4.1.1
Setting Autosampler Parameters
1
Click the [Sampler] tab and specify the number of rinses appropriate for the sample.
4.1.2
Setting GC Parameters
1
Click the [GC] tab and set the analysis conditions.
1 Input an initial temperature for the column oven (40 to 100 °C).
2 Input an injection temperature based on consideration of the boiling point of the target compound
(200 to 300 °C).
3 Select [Split] or [Splitless].