GCMS-QP2010 Series Operation Guide 39
4.4 Analyzing Data
4
3
After checking the mass spectra, click (Register Target Spectrum to Spectrum
Process Table).
The mass spectrum is registered.
Registering identification results in the spectrum process table allows referencing those results at a
later time or outputting them as a report.
4
Close the [Similarity Search Results] window.
4.4.5
Displaying Mass Chromatograms
In the following situations, displaying mass chromatograms (MC) makes it easier to analyze the data.
For instructions of how to display mass chromatograms, see "Appendix G Displaying
Chromatograms" P.96.
Confirming the Purity of Peaks
Displaying mass chromatograms can be used to check the presence of two or more overlapping peaks, or
in other words, to check the purity of a peak in the chromatogram.