GCMS-QP2010 Series Operation Guide 13
2.7 Autotuning
2
2.7
Autotuning
Wait for approximately 2 hours (before starting qualitative analysis) or 4 hours (before starting
quantitative analysis) after starting up the vacuum system and then perform autotuning using the
procedures described below.
2.7.1
Setting Analysis Conditions
If no analysis conditions have been created, start from "2.7.2 Executing Autotuning" P.14.
If a method file is already created, parameters can be specified in the instrument according to the following
procedure.
However, parameters for an accessory or peripheral equipment, except for AOC-20 auto-injector/auto-
sampler, cannot be specified by using the following procedure. When using an accessory/peripheral
equipment, set the parameters on the equipment itself, or by using the software specific to that equipment/
device.
1
Click the [Data Acquisition] icon on the [Real Time] assistant bar.
The [Acquisition] window opens.
2
Click (Open) on the toolbar.