GCMS-QP2010 Series Operation Guide 85
D.2 Easy sTop (Applicable to QP2010 Ultra and QP2010 SE models)
D
5
Perform maintenance by following the instructions displayed on the screen.
Click an image to enlarge it. Click [Back] in the enlarged window to return to the original window.
To perform another maintenance item, click [Back] and repeat the procedure from step 3.
After completing maintenance, close the [MS Navigator] window.
After performing maintenance, reset the usage frequencies and usage times using the procedure
described in "Appendix D.3 Reset Procedure for Usage Frequencies and Usage Times
^
Ver.2.5"
P. 8 7.
D.2
Easy sTop
(Applicable to QP2010 Ultra and QP2010 SE models)
Using Easy sTop allows replacing septa and glass inserts without stopping the vacuum system.
Therefore, it significantly reduces the time required for stabilizing the system after replacement and
eliminates the need for autotuning.
To protect columns, Easy sTop keeps the temperature of the sample injection unit, column oven, and
interface at 70 °C or below. Consequently, it can take about 30 minutes until glass inserts and septa
can be replaced.
1
Double-click one of the icons for consumables in the instrument monitor.
The [Consumable] tab page opens in the [Monitor Settings] window.