11 Maintenance
PFA-100 System – Instruction Manual (US) – Version 1.0.1
311
11
2. Remove the empty bottle.
3. Remove the flip-top cap and insert the new bottle in place.
4. Push the access door closed.
5. Push the knob all the way down to insert the probe into the trigger
solution bottle.
Note: It is normal that residual solution remains in the bottle.
CAUTION
Delay of results caused by air in the trigger solution path as a result of
a missing or insufficiently filled trigger solution bottle.
Minor or moderate injury to the patient can result.
Take care to insert a trigger solution bottle.
Make sure that the trigger solution bottle is sufficiently filled.
NOTICE
Risk of damage to the spiking needle if the spiking needle lever is
pushed down before the access door is fully closed.
Make sure the access door is fully closed before pushing the lever
down.
Note: It is necessary to prime the system after replacing the trigger
solution.