12 Diagnostics and troubleshooting
PFA-100 System – Instruction Manual (US) – Version 1.0.1
312
2. Press the numeric key 4 to select the Print Log option.
3. Press the numeric key 4 to select the Error Log option.
The system will print the serial number and the software version of the
analyzer, followed by the error messages (maximum of the last 25) with
the date and time at which they occurred.
4. To return to the System Ready display, press the softkey located next
to Previous Screen twice.
12.4 Displaying test results
When testing is complete, the analyzer will display the result(s) with the
respective patient ID numbers entered and test type. Events or conditions
(status messages) detected during the test run are displayed below the test
result.
If the test was aborted, *_INVALID TEST_* is shown and the status
message is displayed in the bottom line.
In case no closure is detected during a test until
• the maximum test time has been reached
• the pressure syringe has reached its maximum
• the sample is depleted
1 - Run Control
2 - Maintenance
3 - Setup
4 - Print Log Previous Screen
Previous Screen
1 - Patient Results
2 - Control Results
3 - Usage History
4 - Error Log
Sample A:
* IN VALID TEST *
Flow Obstruction
Sample A:
* IN VALID TEST *
Air Leak