2 Safe handling
PFA-100 System – Instruction Manual (US) – Version 1.0.1
12
2
2 Safe handling
This chapter contains information about:
The intended use (Chapter 2.1)
General safety information (Chapter 2.2)
The meaning of the warning and information symbols on the analyzer
(Chapter 2.4).
The statutory requirements (Chapter 2.5)
The obligations of the system owner (Chapter 2.6).
2.1 Intended use
The PFA-100 Platelet Function Analyzer and Dade PFA-100 Reagents are
in vitro diagnostic devices intended to aid in the detection of platelet
dysfunction in citrated human whole blood.
The PFA-100
®
System is an instrument and test cartridge system in which
the process of platelet adhesion and aggregation following a vascular injury
is simulated in vitro. Platelet dysfunction detected by the PFA-100 System
may be acquired, inherited, or induced by platelet inhibiting agents.
The PFA-100 System is for in-vitro diagnostics only.
The instrument’s functions are:
Moving the carousel (containing PFA test cartridges with samples) into
the instrument
Reading the test code
Dispensing the trigger solution into each PFA test cartridge
Incubating the samples
Aspirating the sample through the aperture of the PFA test cartridge
Measuring the blood flow rate through aperture at a constant vacuum
Determining the closure time by detecting a flow rate, which is for three
seconds below 10% of the initial flow rate at the beginning of the test
Calculating and analyzing the readings
Printing the readings
One or two PFA test cartridges containing samples can be placed in the
instrument after:
Loading the required PFA test cartridges
Entering the test
Then the samples are processed fully automatically as far as the
analysis stage.