3 Delivery and installation
PFA-100 System – Instruction Manual (US) – Version 1.0.1
93
3
3.2.2 Requirements of worktop
The small size of the PFA-100 System requires an area on the laboratory
bench top which is approximately 30cm wide and 50cm deep, with about
50cm clearance above the bench.
The worktop for the analyzer must meet the following requirements:
Stable, horizontal and dry.
To ensure adequate ventilation, a free space of at least 5cm must be left
around the analyzer on all sides.
Load capacity of at least 50kg.
Access to a single power outlet is required within 150 cm of the analyzer
location.
Note: To ensure the analyzer is stable, all of the feet must be in full
contact with the worktop. The analyzer should not be positioned close
to the edge of the worktop.