PFA-100 System – Instruction Manual (US) – Version 1.0.1
29
9.1 Self test and daily alignment check for trigger solution dispensing
To initiate a system self test, proceed as follows:
1. From the System Ready display, press the softkey located next to
Menus.
2. From the Menu display, press the numeric key 2 to select the
Maintenance option.
3. Press the numeric key 2 to select the Self Test option.
False results due to a vacuum leak not being noticed.
False results can lead to misdiagnosis and incorrect treatment.
• When performing the Self Test, strictly follow the instructions given
below.
• Always use a new vacuum test cup and a new cleaning pad.
Run
Print Last Run
Menus
System Ready
Insert Sample(s)
Then Press Run
01/11/07 10:30 AM
Previous Screen
1 - Run Control
2 - Maintenance
3 - Setup
4 - Print Log
5 - Clean O-Ring
6 - Remove O-Ring
7 - Install O-Ring
Previous Screen
1 - Paper Advance
2 - Self Test
3 - Printer Test
4 - Trigger Prime