7 Setup and changing configuration
PFA-100 System – Instruction Manual (US) – Version 1.0.1
77
7
7. To return to the System Ready display, press the softkey located next
to Previous Screen twice.
7.2.4 Patient identification request
This option sets up the system to request the entry of a specific patient
identification number (maximum 12 characters) when running a test.
To set the patient identification request, proceed as follows:
1. From the System Ready display, press the softkey located next to
Menus.
2. Press the numeric key 3 to select the Setup option.
3. Press the numeric key 2 to select the Set Defaults option.
4. Press the numeric key 2 to select the Patient ID option.
Run
Print Last Run
Menus
System Ready
Insert Sample(s)
Then Press Run
01/11/07 10:30 AM
1 - Run Control
2 - Maintenance
3 - Setup
4 - Print Log Previous Screen
Previous Screen
1 - Select Language
2 - Set Defaults
3 - Set Date
4 - Set Time
Previous Screen
1 - Flags
2 - Patient ID
3 - Audible Keys
4 - LIS