PFA-100 System – Instruction Manual (US) – Version 1.0.1
86
6 Description of the analyzer
6
6.3.3 Environmental
6.3.4 General
WARNING
False results or delay of results due to incorrect functioning of the
analyzer.
Death or serious injury to the patient can result.
Only operate the analyzer under the conditions specified below.
Note: This analyzer is for indoor use only.
Parameter Value
Operating temperature 18°C to 29°C
Non-operating storage and
transportation temperature
-29°C to 50°C
Relative humidity 20% to 90%
Pollution degree 2
Altitude (maximum) 2000m
Operating humidity 20% to 90%RH
Non-operating storage and
transportation humidity
0% to 95%RH
Parameter Value
Installation category II
Incubation well temperature
37.9°C C
Incubation well warm-up time
(room to operating temperature)
Max. 15min
Incubation time 3.0min
Cassette capacity 2 PFA test cartridges
Maximum sample volume 1000µL
Vacuum set point 40mBar
Sound pressure level during:
Printing
Priming/Test
77dBA
46dBA