7 Setup and changing configuration
PFA-100 System – Instruction Manual (US) – Version 1.0.1
17
7
7 Setup and changing configuration
This chapter contains information about:
The priming of the analyzer (Chapter 7.1)
The setup of the software (Chapter 7.2)
The configuration of the analyzer (Chapter 7.3)
7.1 Priming the analyzer
To prime the analyzer, proceed as follows:
1. Ensure trigger solution and a priming cartridge (blue cartridge with no
foil on top) is installed in the analyzer.
Figure 7-1 LCD display and keypads
LCD display
Softkeys
Numeric keys
Platelet Function Analyzer
NOTICE
Contamination of the incubation block with trigger solution due to a
defective priming cartridge.
Check that the aluminum foil at the bottom of the priming cartridge is
intact.
Check the priming cartridge for scratches, marks, or debris and replace
or clean it, if required.
Change the priming cartridges periodically, depending on their
condition.
Note: Save the blue priming cartridge for future use.