PFA-100 System – Instruction Manual (US) – Version 1.0.1
27
7 Setup and changing configuration
2. Select Menu > Maintenance > Trigger Prime.
3. Then press the Prime softkey to start priming the trigger system.
The carousel will rotate and the screen will display Trigger Priming,
Please Wait.
When trigger priming is complete, the cassette will rotate to the load
position.
4. Remove the priming cartridge and discard the trigger solution
accumulated during the priming cycle.
5. Rinse the priming cartridge with distilled or deionized water.
6. Press the Continue softkey.
Touching the analyzer carousel while it is turning may result in
pinching of fingers.
Minor or moderate injury can result.
• Always wait until the measurement has finished before inserting or
removing cartridges.
Prime
Cancel
Insert Trigger Bottle
And Priming Cartridge -
Then Press Prime
Continue
Trigger Priming Complete
Remove the Cartridge,
Then Press Continue
Note: When the analyzer is first powered on, at least 15 minutes are
required to reach the operating temperature of 37 °C. When the system
heater block has reached operating temperature, the System Ready
screen will be displayed.