PFA-100 System – Instruction Manual (US) – Version 1.0.1
108
8.4 Printing results
8.4.1 Automatic printing after test
The system is set to automatically print the results after testing is complete.
8.4.2 Printing the results of the last run
To print the results of the last run, proceed as follows:
1. From the System Ready display, press the softkey located next to Print
Last Run.
False results due to problems with the type code sensor or the
readability of the type codes on the PFA test cartridges. This can
cause a mix-up of test types in the instrument software.
False results can lead to misdiagnosis and incorrect treatment.
• Do not use the results obtained whenever the status message TYPE
CODE CALIBRATION REQUIRED or INVALID CARTRIDGE TYPE
DETECTED is displayed.
• Remeasure the samples concerned when the issue is solved.
Run
Print Last Run
Menus
System Ready
Insert Sample(s)
Then Press Run
01/11/07 10:30 AM