PFA-100 System – Instruction Manual (US) – Version 1.0.1
1611
11.6 Replacing fuses
To replace fuses on the analyzer, proceed as follows:
1. Set the power switch to off.
2. Unplug the power cord.
3. Using a small-blade screwdriver, open the fuse compartment and
remove the fuse holder.
4. Place the fuse holder face down with the fuses facing up.
5. Remove the top fuse and replace with a new one.
6. To change the fuse configuration, loosen the Phillips screw with a few
turns.
Electric shock upon contact with live parts.
Death or serious injury can result.
• Before opening the analyzer and replacing fuses, disconnect the mains
plug.
Infection by contaminated parts of the analyzer.
Death or serious injury can result.
• Clean and disinfect the analyzer, see Chapter 10.
Figure 11-5 Fuse compartment
Figure 11-6 Phillips screw