PFA-100 System – Instruction Manual (US) – Version 1.0.1
212
12 Diagnostics and troubleshooting
is completed. During a vacuum leak test, the vacuum pressure of the
system is checked. After successful completion, the message Vacuum
Test Ok will be displayed and printed. If a problem is detected, an error
message will be displayed and printed and the system will prevent the
running of a test. Refer to Chapter 12.5 to take the appropriate action if a
problem occurs.
12.3 Printing error log
The PFA-100 software maintains a record of events or conditions detected
by the system. This record is called the error log and may be useful in
diagnosing problems. To print the error log, proceed as follows:
1. From the System Ready display, press the softkey located next to
Menus.
Figure 12-1 Status messages list
PFA-100
Self Test
REV. 2.2 S/N: 99031
01/10/16 02:32 PM
VACUUM CHUCK OK
CAROUSEL DRIVE OK
INCUBATION SYSTEM OK
VACUUM TEST OK
TRIGGER SYSTEM OK
Test type : Collagen/EPI
Low Flag : 0
High Flag : 300
Test type : Collagen/ADP
Low Flag : 0
High Flag : 300
Run
Print Last Run
Menus
System Ready
Insert Sample(s)
Then Press Run
01/11/07 10:30 AM