PFA-100 System – Instruction Manual (US) – Version 1.0.1
611
11 Maintenance
11
11.3 O-Ring maintenance
This chapter contains information about the manual cleaning and
replacement of the O-Ring.
11.3.1 Maintaining and inspecting the O-Ring
To inspect and maintain the O-Ring, proceed as follows:
1. Remove the O-Ring by performing the Remove O-Ring option from the
Maintenance menu.
2. Rinse the O-Ring under distilled or deionized water. Place O-Ring
between forefinger and thumb and remove any debris by using a
rubbing motion while rinsing under distilled or deionized water.
NOTICE
Soiled and sticky O-Rings can cause the O-Ring removal tool to fail.
To prevent this, thoroughly clean the O-Ring as described in
Chapter 11.3.1 and Chapter 11.3.4 on a regular basis.
Note: The manual O-Ring cleaning procedure has to be carried out
once a week and should be performed whenever the status message
VACUUM TEST FAIL is obtained, after a self test is performed from the
Maintenance menu or whenever the status message TEST
TERMINATED DUE TO AIR LEAK is printed after a test.
The O-Ring should be replaced once a year.
WARNING
Infection by contaminated parts of the analyzer.
Death or serious injury can result.
Wear appropriate personal protective equipment, that is, gloves,
protective clothing, safety glasses, and mask.
Pay attention to the manufacturer's information regarding the gloves'
compatibility with the materials and liquids used.