PFA-100 System – Instruction Manual (US) – Version 1.0.1
87
7 Setup and changing configuration
5. Press the Yes softkey to enable this option (system will request an ID
No. for Every Test) or press the No softkey to disable this option
(system will not request an ID No. for Every Test, but will automatically
assign a unique number).
Upon initial start-up only: The system will automatically begin
initialization.
Otherwise: No further action is performed by the analyzer.
6. To return to the System Ready display, press the softkey located next
to Previous Screen three times.
7.2.5 Analyzer initialization
During initialization, the system will automatically proceed with power on
diagnostics tests. The results of each test will be printed after each test is
completed.
7.3 Configuration
This chapter contains information about:
• Configuring low/high flag ranges for results (Chapter 7.3.1)
• Activating/deactivating audible tones for softkeys (Chapter 7.3.2)
• Transmitting results to a host computer (Chapter 7.3.3)
Yes
No
Previous Screen
Request ID # for Every Test?