PFA-100 System – Instruction Manual (US) – Version 1.0.1
83
3 Delivery and installation
3
3.2.1 Operating conditions
3.2.1.1 Ambient conditions for storage
3.2.1.2 Ambient conditions for operation
The PFA-100 System may be operated under the following ambient
conditions:
Only use indoors
Altitude up to 2000m
Temperature range: 18°C to 29°C
Maximum relative humidity in the permissible temperature range:
20% to 90% RH (non-condensing)
Voltage fluctuations in the mains supply not greater than ±10% of the
rated voltage
Transient overvoltages as they usually occur in the mains supply
Low as possible an influence on the analyzer by steam, solvents and
acid fumes
Protect against direct exposure to sunlight.
Note: The recommended storage temperature for the analyzer is
between -29°C and 50°C. The Non-operating storage and
transportation humidity is between 0% and 95% RH.
WARNING
False results or delayed results due to incorrect functioning of the
analyzer.
Death or serious injury to the patient can result.
Only operate the analyzer under the conditions specified below.
Note: The degree of protection for the PFA-100 System housing
corresponds to IP20 according to IEC 60529.