12 Diagnostics and troubleshooting
PFA-100 System – Instruction Manual (US) – Version 1.0.1
712
12.5.2 Test Terminated Due To Air Leak
The following table contains information about the circumstances in which the message TEST TERMINATED DUE
TO AIR LEAK occurs and how to proceed afterwards.
Description Causes Solutions/Comments Result Interpretation
The analyzer has detected an
initial air leak in the vacuum
system. This condition is
detected only at the
beginning of a test.
• Possible vacuum leak in the
analyzer or a malfunctioning trigger
solenoid pump
• No sample in PFA test cartridge
• Air entrapped in PFA test cartridge
when sample was loaded
• Defective PFA test cartridge
• Verify that sample was added to PFA test
cartridge. Rerun the test with appropriate
sample volume (800µL). Perform a Self Test
via the Maintenance menu without the
cleaning pad to test the vacuum system. If
the vacuum test passes, a malfunctioning
solenoid pump may be the cause for the air
leak. Contact Siemens Service for solenoid
pump troubleshooting procedure. If the
vacuum test fails, perform Manual O-Ring
Maintenance Procedure via the
Maintenance menu. Clean or replace O-
Ring as per maintenance instructions.
Perform self test. If vacuum leak test passes,
rerun sample. If problem persists, contact
Siemens Service.
• Prime system from Maintenance menu.
• Rerun the test with a new PFA test cartridge
and sample.
Non-reportable
Table 12-2 Test Terminated Due To Air Leak