11 Maintenance
PFA-100 System – Instruction Manual (US) – Version 1.0.1
511
11
4. Press the numeric key 2 to select the Maintenance option.
5. Press the numeric key 4 to select the Trigger Prime option.
6. Press the softkey located next to Prime to start priming the trigger
system.
The cassette will rotate and the screen will display Trigger Priming,
Please Wait.
When priming is complete, the cassette will rotate to the load position.
7. Remove the priming cartridge, discard the trigger solution accumulated
in the priming cartridge during the priming cycle.
8. Rinse the priming cartridge with distilled or deionized water.
9. Save for future use or discard if damaged.
10. Press the softkey located next to Continue.
11. To return to the System Ready display, press the softkey located next
to Previous Screen twice.
1 - Run Control
2 - Maintenance
3 - Setup
4 - Print Log Previous Screen
1 - Paper Advance
2 - Self Test
3 - Printer Test
4 - Trigger Prime
5 - Clean O-Ring
6 - Remove O-Ring
7 - Install O-Ring
Previous Screen
Note: We recommend periodically replacing the priming cartridges,
depending on their condition. The priming cartridge must show no
scratches or marks and must be clear of debris. The bottom aluminum
foil must be intact.
Continue
Trigger Priming Complete
Remove the Cartridge,
Then Press Continue