EasyManua.ls Logo

Siemens PFA-100 System - Page 91

Siemens PFA-100 System
138 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
9 QC procedures
PFA-100 System – Instruction Manual (US) – Version 1.0.1
39
9
4. Press the softkey located next to Yes to continue the self test.
5. Load two vacuum test cartridges (create the vacuum test cartridge by
inserting a grayish blue vacuum test cup into the blue priming cartridge)
in position A and B.
6. Press the softkey located next to Continue.
The system will then rotate the carousel and instruct the user to load
the cleaning pad (circular foam sponge) in the well.
7. Once the cleaning pad is in position in the carousel well, apply 4-5 drops
of isopropanol to the center of the pad.
8. Press gently on the pad with gloved index finger 2-3 times to help
distribute the isopropanol.
Yes
No
This Test Will Take
Approximately 7 Minutes
Do You Want To Continue?
NOTICE
Vacuum test failure.
The vacuum test may fail if the following is not observed:
Make sure that a vacuum cup is inserted into the priming cartridge.
Check the priming cartridge (bottom foil intact, no scratches or debris)
and replace it, if required.
Check the vacuum cup for scratches or debris and replace it, if required.
Continue
Cancel
Insert Vacuum Cartridge
in Position A and B
NOTICE
Contamination of the cleaning pad when touching it with bare fingers.
Contamination may be transferred from the cleaning pad to the O-Ring
during cleaning.
Only touch the cleaning pads with appropriate gloves.
Pay attention to the manufacturer's information regarding the gloves'
compatibility with isopropanol.

Table of Contents

Related product manuals