9. Press the softkey located next to Continue.
The system will perform the O-Ring cleaning procedure and a vacuum
test in addition to the power on diagnostics test. The memory test is
performed only during power on and not performed during the self test.
The system will print the pass/fail results each time a test is completed.
After the self tests are completed, the system will print the high/low flag
ranges for each test type.
At the end of the self test, the system will prompt the user to remove the
cleaning pad.
10. Remove the pad and dispose of it in a suitable biohazard waste
container.
11. Press the softkey located next to Continue.
12. Remove the vacuum test cartridge from position B and inspect the
trigger solution dispensing. Check that the trigger solution is dispensed
on the raised platform of the vacuum cup (circled in Figure 9-1).
Infection by contaminated cleaning pads.
Death or serious injury can result.
• After usage, cleaning pads must be treated as a potential source of
infection.
• Wear appropriate personal protective equipment, that is, gloves,
protective clothing, safety glasses, and mask.
• Pay attention to the manufacturer's information regarding the gloves'
compatibility with the materials and liquids used.
Note: The size and exact centering of the drop is not critical. If there is
a drop visible on the platform, the trigger solution dispensing is
adequate. If there is no drop visible on the platform, call your Siemens
Healthcare Diagnostics Service.