PFA-100 System – Instruction Manual (US) – Version 1.0.1
59
13. Remove the vacuum test cartridge from position A.
14. Remove the vacuum test cup from both priming cartridges with gloved
index finger and discard them in a suitable biohazard waste container.
15. Rinse both priming cartridges with distilled or deionized water and save
them for further use.
16. Press the softkey located next to Previous Screen twice to return to the
System Ready display.
Figure 9-1 Trigger solution dispensing
Raised platform of vacuum cup
Infection by contaminated vacuum test cups.
Death or serious injury can result.
• After usage, vacuum test cups must be treated as a potential source of
infection.
• Wear appropriate personal protective equipment, that is, gloves,
protective clothing, safety glasses, and mask.
• Pay attention to the manufacturer's information regarding the gloves'
compatibility with the materials and liquids used.