PFA-100 System – Instruction Manual (US) – Version 1.0.1
183
3 Delivery and installation
3
2. Remove flip-top from trigger solution bottle.
3. Load trigger solution and close access door.
CAUTION
Delay of results caused by air in the trigger solution path as a result of
a missing or insufficiently filled trigger solution bottle.
Minor or moderate injury to the patient can result.
Take care to insert a trigger solution bottle.
Make sure that the trigger solution bottle is sufficiently filled.
NOTICE
Risk of damage to the spiking needle if the spiking needle lever is
pushed down before the access door is fully closed.
Make sure the access door is fully closed before pushing the lever
down.