PFA-100 System – Instruction Manual (US) – Version 1.0.1
107
7 Setup and changing configuration
7
5. Press the softkey located next to Yes to enable or No to disable the Flag
Out Of Range Results option.
6. Use the numeric key to select the PFA test cartridge type to be set for
flag ranges (e.g. Collagen/EPI or Collagen/ADP).
7. Use the numeric key to enter the low flag value (time below which the
result will be flagged).
8. Then press the softkey located next to Enter to save the entry.
9. Use the numeric key to enter the high flag value (time above which the
result will be flagged).
10. Then press the softkey located next to Enter to save the entry.
11. Press the softkey located next to Previous Screen three times to return
to the System Ready display.
Flag Out Of Range Results?
Previous Screen
Yes
No
Previous Screen
1 - Collagen/ADP
2 - Collagen/EPI
Low Flag
0-- sec
Enter
Previous Screen
High Flag
185 sec
Enter
Previous Screen
Note: The system will default to No if no selection is made.