PFA-100 System – Instruction Manual (US) – Version 1.0.1
89
9 QC procedures
9
11. Press the softkey located next to Enter.
12. Use the numeric keypad to enter the control ID (maximum
12 characters).
13. To modify the control ID, use the < key to erase it, then enter the correct
number.
14. Press the softkey located next to Run.
When the testing is complete, the analyzer will print and display the
results along with the test type, the control ID number and lot number
entered, a CV% and a MEAN value if two tests were run.
15. Remove the cassette carefully from the carousel: Holding the cassette
in one hand, remove the PFA test cartridge by gently pulling the bottom
of the PFA test cartridge toward you until it unsnaps.
16. Dispose of the PFA test cartridge in a suitable biohazard waste
container.
Enter Cartridge Lot Number
Enter
Cancel
Note: To minimize the risk of trapping an air bubble, place the tip of the
pipette against one of the inside corners of the opening while slowly
pipetting in the sample.
Note: Events or conditions (status messages) detected during the test
run are indicated below the printed test result. Refer to Chapter 12.5
and Chapter 12.6, for full explanation of status messages.