Scale parameters and functions
8.3 DR 3 calibration parameters
SIWAREX WP231
Manual, 07/2014, A5E31238908A-02
67
Standstill
range
(Page 73)
Standstill range (in d) FLOAT 4 rwP 0.1 0 maximum
number
range+
1048
Standstill
time
(Page 73)
Standstill time 1 in ms TIME 4 rwP 2000 10 10000 1050
Standstill
waiting time
(Page 74)
Waiting time until standstill.
0: Any scale commands de-
pendent on standstill are reject-
ed immediately in the absence
of standstill.
>0: Maximum waiting time until
command is executed
TIME 4 rw 2000 0 10000 1052
Low-pass
filter limit
frequency
(Page 74)
Low-pass filter 1 - limit frequen-
cy:
0: Filter disabled
FLOAT 4 rwP 0.5 tbd tbd 1054
Low-pass
filter number
(Page 74)
Ordinal number of low-pass
filter 1:
Filter ordinal number 2*(1 … 5)
USHORT 2 rwP 4 2 10 1056
Reserve 2 Reserve USHORT 2 rw 0 - - 1057
Low-pass
filter limit
frequency
(commission-
ing)
(Page 75)
Low-pass filter 2 - limit frequen-
cy:
0: Filter disabled
FLOAT 4 rw 0 tbd tbd 1058
Low-pass
filter number
(commission-
ing)
(Page 75)
Ordinal number of low-pass
filter 2:
Filter ordinal number 2*(1 … 5)
USHORT 2 rw 4 2 10 1060
Mean value
filter depth
(Page 75)
Filter for digit values,
permitted filter depth: 0 … 250
USHORT 2 rwP 10 0 250 1061
Display
weighing
range data
(Page 75)
Display weighing range data
0: No
1: Yes
USHORT 2 rwP 0 0 1 1062
Interface for
legal trade
display
(Page 75)
Selection of interface for Se-
cureDisplay:
0: HMI control at ETHERNET
1: HMI control via S7 interface
USHORT 2 rw 0 0 3 1063
Firmware
version
SecureDis-
playHeader
Maximum length and actual
length of string for FW version
SecureDisplay
UBYTE[2] 2 rw 12, 12 1064
Firmware
version
SecureDis-
play
(Page 75)
FW version of the SecureDis-
play used on the HMI
CHAR[12] 12 rw 1065
Reserve Reserve UBYTE[2] 2 rw 0 - - 1071
Reserve Reserve CHAR[20] 20 rw 0 - - 1072