REMOTE BUS INTERFACE: RS-232
0 = Power-on load previous enable
registers
Power-on Status Clear Query
Standard Event Status Register Query
Standard Event Status Enable
Command
Standard Event Status Enable Query
BIT 0, 01H,(1) Operation Complete
BIT 1,02H,(2) Not Used
BIT 2,04H,(4) Query Error
BIT 3,08H,(8) Device Error
BIT 4,10H,(16) Execution Error
BIT 5,20H,(32) Command Error
BIT 6,40H,(64) Not Used
BIT 7,80H,(128) PROMPT
Service Request Enable Command
Service Request Enable Query
*IDN?
Read the instrument identification string. Company =SLA.
*RST
Reset the instrument to original power on configuration. Does not clear Enable register
for Standard Summary Status or Standard Event Registers. Does not clear the output
queue. Does not clear the power-on-status-clear flag.
*TST?
Performs a self test of the instrument data memory. Returns 0 if it is successful or 1 if the
test fails.
*CLS
Clears the Status Byte summary register and event registers. Does not clear the Enable
registers.
*OPC
Sets the operation complete bit (bit 0) in the Standard Event register after a command is
completed successfully.
*OPC?
Returns an ASCII "1" after the command is executed.
*WAI
After the command is executed, it prevents the instrument from executing any further
query or commands until the no-operation-pending flag is TRUE.
*PSC {1|0}