REMOTE BUS INTERFACE: RS-232
Sets the power-on status clear bit. When set to 1 the Standard Event Enable register and
Status Byte Enable registers will be cleared when power is turned ON. 0 setting indicates
the Enable registers will be loaded with Enable register masks from non-volatile memory
at power ON.
*PSC?
Queries the power-on status clear setting. Returns 0 or 1.
*ESR?
Queries the Standard Event register. Returns the decimal value of the binary-weighted
sum of bits.
*ESE <value>
Standard Event enable register controls which bits will be logically ORed together to
generate the Event Summary bit 5 (ESB) within the Status Byte.
*ESE?
Queries the Standard Event enable register. Returns the decimal value of the binary-
weighted sum of bits.
*STB?
Read the Status Byte. Returns the decimal value of the binary-weighted sum of bits.
*SRE <value>
Service Request enable register controls which bits from the Status Byte should be use to
generate a service request when the bit value = 1.
*SRE?
Queries the Service Request enable register. Returns the decimal value of binary-
weighted sum of bits.