Page 104 (Model J)
5. Ensure the pressure regulator leads
(J) are routed as shown.
6. Connect the electrical connectors to
PR3 (K), PR4 (L) and PR6 (M).
PR3 cable is tagged 3.
PR4 cable is tagged 4.
PR6 cable is tagged 6.
16.6.8 Installation of the new blender assembly (L0S80/405/002)
7. Using the original nuts and washers (K) fit the new blender assembly (M).
J
K
L
M
K
K
K
K
M