(Model J) Page 151
Step 52 Set a Tidal volume (Vte (TTV)) of 8ml and a Max PIP of 30mbar
Verify that:
The displayed waveform changes
showing a rise in flow with a sharp
cut off.
Step 53 Turn off the volume limiting.
Step 54 Press the Mode Select button and
select PSV from the Mode Select panel, but do
not press the confirm button. Advance to Step 55.