Step 49 From the mode select panel press the TTV
plus
button.
(Skip Step 49 to Step 53, if functional testing
without a flow sensor).
Step 50 Turn on the Volume targeting by pressing the
button marked OFF. The text in the button will
change to ON.
Press the Return button to return to the Mode
Panel.
Step 51 The Tidal Vol parameter will now appear.
Note: The PIP parameter control changes to the MAX PIP parameter control
when volume limiting is turned on.