(Model J) Page 337
GREEN Inop - “O2 CELL
CALIB”
None Indicates that the oxygen cell is
currently being calibrated. The
parameter value for FIO_2 is
displayed as -?-
GREEN Inop - “PR.
SENS. DRIFT”
Pres. Sens. Drift (11) Indicates that pressure sensor
drift has occurred.
RED Alarm -
“LOW PRESSURE”
Sust. sub ambient (2)
Sub ambient (3)
Low Pressure (9)
Indicates one of three low
pressure alarm conditions.
RED Alarm -
“CONTIN. PRESS”
Cont. positive press (7)
RED Alarm -
“HIGH PRESSURE”
High Pressure (8)
RED Alarm -
“CYCLE FAIL”
Fail to cycle (10)
Breath Not Detec. (38)
Indicates one of two alarm
conditions.
RED Alarm -
“PR. SENS. DRIFT”
Press. Sens. Drift (11)
RED Alarm -
“HFO P CHANGE”
Rise in Mean P (12)
Drop in Mean P (13)
Rise in Max P (14)
Drop in Max P (15)
Rise in Min P (16)
Drop in Min P (17)
Indicates a change has occurred
in the HFO pressure waveform.
RED Alarm -
“GAS SUPPLY FAIL”
No Gas (18)
No O2 Supply (19)
No Air Supply (20)
Indicates that one or both of the
input gas supplies have failed.
RED Alarm -
“BATTERY FAULT”
No Battery Detec. (21)
RED Alarm -
“BATTERY LOW”
Battery Low (22)
RED Alarm -
“BLOCK - F.G.
Blocked Fresh Gas (24)
RED Alarm -
“LEAK - F.G.”
Leaking Fresh Gas (25)
RED Alarm -
“APNOEA”
Apnoea - Volume (39)
Apnoea - Pressure (40)
YELLOW Alarm -
“FLOW SENS ALARM”
Flow sens. defect (28)
Clean Flow Sensor (29)
Connect Flow Sens (30)
Cannot Cal. Flow (31)
Cal. Flow Sensor (32)
Indicates one of five alarm
conditions associated with the
flow sensor.