(Model J) Page 91
7. Remove the two screws retaining the bracket to
the battery.
8. Reversing the procedure for replacement of the batteries, re-install all the electronic
module components.
Note: When refitting the batteries the screws require a torque setting of 1.2Nm
9. Carry out the following to complete the 24 month overhaul.
10. Carry out the 12 & 36 month preventative maintenance procedure.
16.4.4 Setup
1. Reset the time and date.
2. Carry out the two point oxygen calibration.
3. Carry out a full functional test.
4. Replace all the covers.