Engineering Value
Specify the calibration value in engineering units.
Input Signal Value
Specify the corresponding signal value for the defined value and defined span
calibration modes.
Calibrate
Select calibrate to begin calibration for the selected channel. If the transducer is
already calibrated, TCE presents more calibration options.
NOTE
Calibration is only available for defined channels on a connected eDAQ
lite
.
Calibration Date
TCE automatically enters the calibration date upon transducer calibration.
6.1.6 Prerun Rezero
Mode
Select the prerun rezero mode for the transducer. Interactive rezeroing is available for
all modes except “Not applicable.” Prerun rezeroing does not affect the fundamental
transducer calibration definition. Prerun zero offsets are effective for the duration of the
test only.
NOTE
The eDAQ
lite
never performs rezeroing on test runs started after a reset caused by
an exception error (e.g., Timeout, Calibration or DeviceOverFlow).
Mode Description
Not applicable No rezeroing required.
Interactive only Rezero interactively only (no automatic rezeroing).
First run only Automatically rezero immediately prior to the start of the first test run
only.
All runs (except
power fails)
Automatically rezero immediately prior to the start of every test run
except test runs initiated by eDAQ
lite
power fail resets.
All runs Automatically rezero prior to the start of every test run. This option is
not advised for low level inputs.