Somat eDAQ
lite
198
I2773-4.4 en
NOTE
Always provide the +12 volt REF voltage for the SWC interface. For other CAN
interfaces, the red power wire can be used to power transducers with a
user-configurable range of 3 to 24 volts.
12.3 ELHLS (High Level Analog Layer)
ELHLS Transducer Cable
The ELHLS layer uses a Somat SAC-TRAN-MP Transducer Cable
(1-SAC-TRAN-MP-2-2 or 1-SAC-TRAN-MP-10-2) with an M8 connector and a set of
color-coded pigtail wires. The following table lists the pinouts for the SAC-TRAN-MP
cable when used for an ELHLS input.
12.4 ELBRG (Bridge Layer)
ELBRG Transducer Cable
The ELBRG layer uses a Somat SAC-TRAN-MP Transducer Cable (SAC-TRAN-MP)
with an M8 connector and a set of color-coded pigtail wires. The following table lists
the pinouts for the SAC-TRAN-MP cable when used for an ELBRG input.
Function Pin Wire Color
SWC 1 Brown
CANH 2 White
AGnd 3 bare wire
Power 5 Red
+12 V REF 5 Red (SWC)
CANL 6 Green
Function Pin Wire Color
reserved 1 Brown
+ Signal Input 2 White
Shield 3 bare wire
Ground 4 Black
Power 5 Red
- Signal Input 6 Green
Function Pin Wire Color
reserved 1 Brown
+ Signal Input 2 White
Shield/Ground 3 bare wire
HBM: public