I2773-4.4 en
27
Somat eDAQ
lite
2.1.3 Updating Firmware
HBM regularly releases updates to the Somat eDAQ
lite
firmware that expand
functionality and fix known bugs. To download compatible firmware and software, visit
www.hbm.com/somat.
NOTE
The eDAQ
lite
firmware uses an alpha-numeric version number (e.g., 3.11.A) while
TCE uses a numeric version number (e.g., 3.11.0). For compatibility, the first two
numbers of the eDAQ
lite
and TCE version numbers must be identical (e.g., 3.11).
After running the installer on the support PC to copy the files to the Somat directory,
follow the steps below to upgrade the eDAQ
lite
firmware. Always update the ELCPU
firmware before updating any layer level firmware.
CAUTION
If updating from an earlier release version, do not attempt to perform the upgrade if the
eDAQ
lite
is in need of time-critical testing. An upgrade failure of the eDAQ
lite
firmware can render the eDAQ
lite
inoperable until it is upgraded at the factory.
HBM: public