I2773-4.4 en
51
Somat eDAQ
lite
Prompt user for run descriptions on test run starts
Force TCE to issue a prompt for a test run description at the start of every test run. As
this option is the only method of entering test run descriptions, deselecting it ensures
blank test run description fields.
Require user to verify PC card purge during test initializations
Force TCE to issue a verification prompt in order to purge the PC Card during test
initialization. If not selected, TCE purges the PC Card automatically.
NOTE
If multiple SIE files exist on the eDAQ
lite
, TCE issues the verification prompt
regardless of the preference setting.
Allow test initializations with channels that are not calibrated
Allow TCE test initializations with non-calibrated transducer channels. This option is
not applicable to strain transducer channel types.
Auto set current communication preference on TCE setup file open
When opening a TCE setup file, automatically set the current communication
preferences. If the existing preference matches any network node defined in the
opened file, the preference remains unchanged. Otherwise, TCE sets the active
eDAQ
lite
to the master network node or, if no master exists, the first defined network
node.
Warn user if...
Warn user about unused channels
Select one of three options of when TCE should issue a warning about unused
channels.
Warn user if... Description
Ratio of Calibration
Span to Full-Scale
Span <
Issue a warning if the calibration span is less than the specified
percent of the full-scale span for a transducer channel (or vice-versa).
Difference Between
Host PC and FCS
Clocks >
Issue a warning if the difference between the host PC real time clock
and the eDAQ
lite
real time clock exceeds the specified number of
minutes.
Deviation of Shunt
Calibration Volts from
Ideal >
Issue a warning if the measured shunt calibration span in volts
deviates from the ideal shunt calibration by at least the specified
percentage.