EasyManua.ls Logo

Spectrasensors OXY5500 - Two-Point Calibration

Spectrasensors OXY5500
140 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Oxygen Analyzer
318 4900002239 rev. G 7-11-17
1. Select Calibration from the SENSOR OPTIONS window. Refer to
Figure 3–22.
2. Press .
A message window will display requesting a response to the
following question, “Measurement active. Abort for Configuration?”
Refer to Figure 3–23.
3. Select Yes to stop the measurement in order to move to the
CALIBRATION windows. Use the buttons to navigate
between input fields.
Figure 3–22 Calibration button in the Sensor
Options window
Figure 3–23 Message window - Stop
measurements during configuration

Table of Contents